Allergen-Shield Peanut ELISA

QUANTITATIVE ELISA FOR PEANUT ALLERGENS

  • Quantitative results in 30 minutes after sample preparation
  • High specificity using monoclonal antibodies
  • Uniform protocol across the Allergen‑Shield ELISA range
  • Broad food matrix compatibility
  • Surface swabbing kits available
  • Excellent reproducibility and precision

 

For the in‑vitro quantitative determination of peanut residues in food matrices such as chocolate, bakery products, sauces, cereals, snacks, and processed foods. This test is intended for use by trained laboratory professionals for food allergen monitoring and quality assurance testing.

 

The Allergen‑Shield Peanut ELISA is a quantitative sandwich enzyme‑linked immunosorbent assay (ELISA) designed for the detection of peanut residues in food products labeled as peanut‑free. The assay identifies trace levels of peanut proteins, ensuring compliance with international food allergen labeling regulations (EU No. 1169/2011, FDA, FSANZ, and Codex Alimentarius).

General Information

Peanut (Arachis hypogaea) is one of the most common and potent food allergens worldwide. It contains multiple allergenic proteins, such as Ara h 1, Ara h 2, and Ara h 3, which are heat‑stable and resistant to food processing. Even trace contamination can cause severe allergic reactions, including anaphylaxis. Therefore, sensitive and reliable detection of peanut residues is essential for allergen management and consumer safety.

Principle of the Test

The Allergen‑Shield Peanut ELISA employs a sandwich enzyme immunoassay format. Microtiter wells are pre‑coated with antibodies specific to peanut proteins. Standards and extracted samples are added to the wells; any peanut proteins present bind to the immobilized antibodies. Following washing, a peroxidase‑conjugated detection antibody binds to the captured antigen to form an antibody‑antigen‑antibody complex. The addition of TMB substrate produces a color proportional to the peanut concentration. The reaction is stopped by acid addition, and the absorbance is measured at 450 nm.

 

Performance Characteristics

  • Limit of Detection (LOD): 0.25 ppm peanut (0.625 ppm peanut protein)
  • Limit of Quantification (LOQ): 1.0 ppm peanut (0.25 ppm peanut protein)
  • Standard curve range: 0–25 ppm
  • IC50: 8–16 ppm
  • Recovery: 95.6% (CV = 6.1%)
  • Intra‑assay CV at 5 ppm: 6.9% (n=16)
  • Intra‑assay CV at 15 ppm: 7.2% (n=16)
  • The assay exhibits strong reactivity to peanut (Arachis hypogaea).
  • At concentrations of 500,000 ppm pumpkin seed (0.0005%), pine nut (0.0003%), chestnut (0.0002%), brown rice flour (0.0002%), chickpea (0.0001%), and lentil flour (0.0001%) were minimally reactive in the assay. (Given the reported adulteration of some commercial spices with peanut in recent years, it is possible that the reactivity observed in certain matrices (cumin = 0.0004%, cayenne pepper = 0.0002%, and nutmeg = 0.0005%) does not represent real cross reactivity within the assay).

 

 

Storage and Stability

Store kit and reagents at 2–8 °C. Do not freeze. Protect standards and TMB substrate from light. Reseal unused strips with desiccant immediately after use. Expiry dates are printed on labels.

 

 

Resource

   Allergens Shield Brochure

   Allergen-Shield Peanut ELISA Kit Manual 

 

Ordering Details

Stock Code        Description                                                                      Pack Size

PGB-A1596           ALLERGEN-SHIELD PEANUT KIT                                          96

PGB-A1548           ALLERGEN-SHIELD PEANUT KIT                                          48

PGB-AS1020         ALLERGEN-SHIELD SWABBING KIT                                    20

Enquire Now