Clean-Trace Surface Positive Control ATP10
TO VERIFY THE PERFORMANCE OF CLEAN-TRACE SURFACE ATP SWABS

The ATP10 Clean-Trace Surface Positive Control is a precision reference reagent designed to verify the functionality and reliability of Neogen® Clean-Trace® Surface ATP Testing systems. It ensures the accuracy of ATP detection instruments and reagent performance through a controlled ATP standard concentration. Each test vial contains a known ATP load used to validate both reagent integrity and luminometer operation.
Product Description and Intended Use
The ATP10 Clean-Trace Surface Positive Control provides assurance that Neogen® Clean-Trace® reagents and instruments are operating correctly. Each kit contains 10 ready-to-use reagent vials, each containing 3.0 ng or 5 × 10⁻¹² moles of ATP. A pre-moistened test swab is used to collect the ATP pellet within the vial, which is then processed using the Clean-Trace® Surface ATP Test protocol. Expected readings should exceed 10,000 RLU, confirming correct instrument and reagent function.
Storage and Stability
- Store between 2°C and 8°C (36°F–46°F) for maximum shelf life.
• Do not freeze.
• Protect from direct sunlight.
• Ensure reagents reach room temperature prior to testing.
• Expiry date and lot number are printed on each package.
Instructions for Use
-
- Allow all test components (swabs and positive control vials) to reach room temperature for at least 10 minutes.
- Gently tap the vial to ensure the ATP pellet is settled at the bottom.
- Remove the stopper carefully to avoid contamination.
- Open a Clean-Trace® test swab and swab the entire interior surface of the vial to collect the ATP pellet.
- Reinsert the swab into the device and activate the test by pressing the handle fully down.
- Shake the activated swab side-to-side for at least 5 seconds to ensure complete mixing.
- Read the test result on the Clean-Trace® luminometer per the manufacturer’s instructions.
- Expected result: greater than 10,000 RLU.
- Dispose of used materials according to biohazard waste guidelines.
Troubleshooting Guide
Low signal readings may be caused by:
• Incomplete mixing of sample – shake and retest.
• Expired or improperly stored reagents.
• Reagents not equilibrated to room temperature before testing.
• Residual ATP pellet remaining in vial (incomplete swabbing).
• Contaminated or damaged luminometer.
Safety and Handling
ATP is a naturally occurring compound found on skin and many surfaces. To prevent contamination:
• Do not touch the swab or inside of the vial.
• Wear gloves when handling reagents and samples.
• Dispose of used vials and swabs according to local biohazard regulations.
Consult the Safety Data Sheet (SDS) for further information.
Resources
ATP10 Clean-Trace Surface Positive Control
Ordering Details
Stock Code Description Pack Size
3MATP10ATP Clean Trace Surface Positive Control Vial 10